EVAR Technical Tip – Confirmation of Contralateral Limb Gate Cannulation Using a Moulding Balloon☆
Abstract
Introduction
Accurate confirmation of cannulation of the shorter contralateral limb gate of an abdominal aortic endograft can be challenging. Catheter angiogram may not exclude all possible errors.
Report
Accurate contralateral cannulation can be confirmed by insertion of a moulding balloon over a stiff wire and gentle inflation of the moulding balloon across the contralateral gate of the main body component.
Discussion
The technique of using a moulding balloon to confirm accurate cannulation of the contralateral gate is a good method of eliminating doubt whilst maintaining a stable wire and catheter position.
Keywords: Abdominal aortic aneurysm, Endovascular aneurysm repair, Cannulation
☆ This work has not been present at any society meetings.
PII: S1533-3167(10)00021-X
doi:10.1016/j.ejvsextra.2010.06.001
© 2010 European Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.