EJVES Extra
Volume 18, Issue 1 , Pages 6-8, July 2009

Anaconda AAA Bifurcated Endovascular Graft Migration: Customised Aortouniiliac Endograft Conversion

Angiology, Vascular and Endovascular Surgery Service, Instituto Vascular Internacional, Hospital Universitario Madrid Montepríncipe Hospital Madrid Torrelodones, Ave./Monteprincipe 25, 28660 Boadilla del Monte, Madrid, Spain

Received 4 January 2009; accepted 4 March 2009. published online 22 May 2009.

Abstract 

We report a case of a 58-year-old-man with a 55-mm infrarenal aortic aneurysm treated with an Anaconda™ modular endograft. An error during a delivery-system pull-back manoeuvre caused the migration of the endograft. This was resolved with an aorto-uniiliac conversion and a femoro-femoral crossover bypass. Since a ready-made aorto-uniiliac system was not available, we customised one using a new Anaconda device, squeezing the contralateral gate and occluding this limb with coils.

Keywords: Aortic aneurysm, Endovascular repair, Complications EVAR

 

PII: S1533-3167(09)00009-0

doi:10.1016/j.ejvsextra.2009.03.003

Refers to article:

  • Anaconda AAA bifurcated endovascular graft migration: Customized Aortouniiliac endograft conversion. , 22 May 2009

    European Journal of Vascular & Endovascular Surgery August 2009 (Vol. 38, Issue 2, Page 251)

EJVES Extra
Volume 18, Issue 1 , Pages 6-8, July 2009